Electromagnetic Compatibility Guidance
LIFEPAK 1000 Defibrillator Operating Instructions G-5
©2018 Physio-Control, Inc.
1700-2100
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1, 3, 4, 25
UMTS
DCS-1800
PCS-1900
UTRA Band 1, 2, 4, 9
2.0 0.3
d = 6/9
2400-2570
Bluetooth
WLAN
802.11 b/g/n
RFID 2450
LTE Band 7
4G/LTE-A
2.0 0.3
d = 6/28
5100-5800 WLAN 802.11 a/n 0.2 0.3
d = 6/9
For transmitters in the frequency ranges listed above, the recommended separation distance d in
meters (m) can be determined using the equation in the table above with the known power, P of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1:These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Band (MHz) Service
Maximum
Power (W)
Distance (m) Equation