[N.B]
The real vertical deviation in V/div is equal to the attenuation
factor multiplied by the range of vertical deviation selected on
the oscilloscope. It will be doubled in the case of use of a 50 Ω
load.
Example:
With the probe on factor x 200, the oscilloscope on 0.5 V/div,
the real vertical deviation is 200 x 0.5 = 100 V/div.
With a 50Ω load on the input of the oscilloscope the deviation
becomes 200 V/div.
1000
Real Deviation In V/div
1 500 200
5000.5 250 100
2000.2 100 40
1000.1 50 2
5050 m 25 10
2020 m 10 4
1010 m 5 2
55 m 2.5 1
22 m 1 0.4
DP-50 Instruction Manual
100
50
20
10
5
2
1
0.5
0.2
Vertical Deviation on the
Oscilloscope in V/div
11
6. MAINTENANCE
For maintenance, only use specified spare parts.
The manufacturer can not be held responsible for any accident
arising following a repair made other than its after sales service or
approved repairers.
x 1000 x 500 x 200 x 100