50
PROBLEM POSSIBLE CAUSE SOLUTION
LCD Display
shows battery
symbol
Low Battery Change the batteries
LCD Display
shows
“EE”
Arm or blood pressure monitor
was moved during testing
Re-test taking care to not move
your arm or the blood pressure
monitor
The cuff does not inflate proper-
ly or pressure falls quickly during
testing
Make certain the plugs and bra-
ckets of the cuff are securely
attached to the monitor.
The cuff was not properly applied.
Apply the cuff correctly and try
again
No response
when you press
button or load
battery.
Incorrect operation or strong elec-
tromagnetic interference.
Take out batteries for five min-
utes, and then reinstall all bat-
teries.
In Clock Mode,
no response
when you press
button.
The touch switch is in OFF position.
Please keep touch switch in ON
position, and try again
LCD Display
shows abnormal
result
The cuff position was not correct or
it was not properly tightened
Apply the cuff correctly and try
again
Body posture was not correct dur-
ing testing
Review the “BODY POSTURE
DURING MEASUREMENT” sec-
tions of the instructions and
re-test.
Speaking, arm or body movement,
angry, excited or nervous during
testing
Re-test when calm and with-
out speaking or moving during
the test
Irregular heartbeat (arrhythmia)
It is inappropriate for people
with serious arrhythmia to use
this blood pressure monitor.
12. TROUBLESHOOTING