EasyManua.ls Logo

Plusmed pM-KN01 - Page 26

Default Icon
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
2. Connect the power, put the plug of power line connected with the power socket of the oxygen generator, and the
other end of the plug connects with indoor power socket, turn on the power switch. (See diagram 5)
Diagram 5
3. Adjust well oxygen output ow according to the request, the red scale of the ow meter is MAXIMUM recommended
oxygen absorbing ow (counterclockwise—on, clockwise—off, for common single treatment: pM-KN01 adjusted to 5L/
min; For double use: pM-KN01 adjusted to 3-5L/min.(see diagram 6)
Diagram 6
4. Insert the intake end of absorbing oxygen tube onto the outlet of humidifier, then set the absorbing oxygen tube over
patient’s ears, insert the nasal tube into patient’s nostrils to absorb oxygen; the best absorbing time for health care
keeps 40-50 minutes per time, absorbing time for medical treatment shall be followed doctor’s advice.
(see diagram 7)
Diagram 7
WARNING: 5 minutes are needed from oxygen concentrator from warming up to reach regular function and nominal
performance.
Red scale

Related product manuals