27
5. After breathing oxygen, turn off the power, and pull up the nasal oxygen cannula from humidification bottle. If turn off
the machine without pull off the nasal oxygen cannula, it may cause oxygen tube bending and made remain water in the
humidification bottle backow to machine, further lead to machine failure working .When discontinuous using, please
unplug the power plug.(see diagram 8)
Atomization operation methods
The atomization function of the concentrator is applicable to help to cure the sufferers of chronic pulmonary tuberculosis
and respiratory system, etc.
:Atomization treatment must be under advice and suggestion of your physician.
:Using distilled water to do the atomization for several seconds after each operation may lighten the crystallization
caused by medical solution.
:If atomizing cannot work then open the cover of the bottle and add clean water in small amount. Rotate the white
ball which lies in the bottle with the gas resource connected and select the proper angle to gain a better atomization.
a. Open the cover of medical cup, and add atomization remedy that needed, then close the cover.
b. Connect the joint of atomization nozzle (or mask) with the cover of medical cup, and then connect the other end of
atomizer connection tube with the atomization outlet and turn on atomizer.
c. Turn on the power of oxygen concentrator, and shut up ow meter, then it is ready for atomization treatment.
d. Do clean the atomization devices followed by the instructions of the atomization devices.
Unplug the power plug
Diagram 8
6. If the patient need timing oxygen absorbing, please refer to instruction on page 7 item 9 (Timing keys).