Ultrasound Applicator
ActionDescriptionCause of FailureEect of Failure
Check that Ultrasound
Applicator is properly
connected with locking
mechanism.
Replace Ultrasound
Applicator.
If Applicator still does not
operate properly, call service.
Warning on the screen
“COMMUNICATION
ERROR”.
No indication on the
Applicator.
Connection problemUltrasound
Applicator
does not operate
Verify applicator connection to
the system. In case of green LED
not operating, call a Pollogen
authorized service center
No ON/OFF indication of
the applicator.
Green LED not operatingNo Green LED
indication
In case of Orange LED not
operating, replace Applicator.
If Applicator still does not operate
properly, call a Pollogen authorized
service center.
No indication
for temperature
measurements.
Orange LED not operatingNo Orange LED
indication
During ULTRASOUND Output,
verify that once the green LED
turned OFF, the Blue LED turned
ON. Otherwise, call a Pollogen
authorized service center.
No indication of
Ultrasound
output
Blue LED not operatingNo Blue LED
indication