10 5714V104-UK
Auxiliary supply:
2 wire supply, pin 46...45 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25...15 VDC / 0...20 mA
TC input
Type Min. value Max. value Standard
B
E
J
K
L
N
R
S
T
U
W3
W5
LR
0°C
-100°C
-100°C
-180°C
-200°C
-180°C
-50°C
-50°C
-200°C
-200°C
0°C
0°C
-200°C
+1820°C
+1000°C
+1200°C
+1372°C
+900°C
+1300°C
+1760°C
+1760°C
+400°C
+600°C
+2300°C
+2300°C
+800°C
IEC 60584-1
IEC 60584-1
IEC 60584-1
IEC 60584-1
DIN 43710
IEC 60584-1
IEC 60584-1
IEC 60584-1
IEC 60584-1
DIN 43710
ASTM E988-90
ASTM E988-90
GOST 3044-84
Cold junction compensation (CJC) via internal sensor . . . . . . . . . . . . . . ±(2.0°C + 0.4°C * Δt)
Δt = internal temperature - ambient temperature
Sensor error detection, all TC types. . . . . . . . . . . . . . . . . . . . . . . . . Yes
Sensor error current:
when detecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Nom. 2 μA
else. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 μA
Basic values
Input type Basic accuracy Temperature coecient
mA ≤ ±4 μA ≤ ±0.4 μA / °C
Volt ≤ ±20 μV ≤ ±2 μV / °C
Pt100 ≤ ±0.2°C ≤ ±0.01°C / °C
Linear resistance ≤ ±0.1°Ω ≤ ±0.01 Ω / °C
Potentiometer ≤ ±0.1°Ω ≤ ±0.01 Ω / °C
TC type:
E, J, K, L, N, T, U
≤ ±1°C
≤ ±0.05°C / °C
TC type: R, S,
W3, W5, LR
≤ ±2°C
≤ ±0.2°C / °C
TC type: B
85...200°C
≤ ±4°C
≤ ±0.4°C / °C
TC type: B
200...1820°C
≤ ±2°C
≤ ±0.2°C / °C
EMC - immunity influence. . . . . . . . . . . . . . . . . . . . . . . . . < ±0.5% of readout