<status> = an ASCII character that can take the following values:
“S” = stable weight
“M” = weight that is not stable (moving)
“O” = weight greater than the maximum capacity
“E” = weight that cannot be detected.
<s1>...<s2> = 6 ASCII characters of setpoint.
<net weight> = 6 ASCII characters of net weight.
<gross weight> = 6 ASCII characters of gross weight.
<peak> = 6 ASCII characters of peak.
If the request is made cyclically, the weight is acquired with a maximum frequency of:
Frequency Baud Rate
200Hz 115200
50Hz 38400
35Hz 19200
25Hz 9600
8Hz 2400