WEIGHING PARAMETERS
The parameters in this menu allow the adjusment of the acquisition and updating timing of the display
and the manual or automatic zeroing performed by transmitter
F1LtEr WEIGHT FILTER
This parameter adjusts the refresh speed of the display and the serial
and analog output.
Low values of the filter speed up the display refresh.
High values of the filter slow down the display refresh.
Value Update Response
0 50 Hz 25 Hz
1 50 Hz 16 Hz
2 25 Hz 8 Hz
3 25 Hz 5 Hz
4 25 Hz 2.5 Hz
5 10 Hz 1.5 Hz
6 10 Hz 1 Hz
7 10 Hz 0.7 Hz
8 5 Hz 0.4 Hz
9 5 Hz 0.2 Hz
Default: 5
Not1on WEIGHT STABILITY
This parameter defines the divisions number needed to deem the
weight stable.
A large number of divisions allows the transmitter to detect quickly
the weight stability, which is needed when executing tare and print
commands.
Value Change
0 Always stable weight
1 Stability determined quickly
2 Stability determined with medium parameters
3 Stability determined accurately
4 Stability determined with the highest accuracy
Default: 2
Auto 0 AUTOZERO UPON SWITCHING ON
This parameter defines the value of the maximum resettable weight
when the instrument is switched on.
This operation corresponds to a zero calibration of the system and
is executed only if the weight is stable and below the set value.
Value from 0 to the value of the CAPAC parameter.
Default: 0