Page 13
INFO DISPLAY
When the instrument is switched ON, you can test the display, then in sequence you can display the
identification code of the software and its version. Communication codes in the event of a request for
assistance.
ERRORS NOTIFICATION
In the operation mode, the display can report the following error codes.
ϩϩϩϩϩϩ the weight applied to the load cell exceeds by more than 9 divisions the maximum capacity
of the weighing system.
O-L Signal of the load cells absent or outside of the measurement range mV/V.
No CoN Fieldbus network disconnected
E-ProF PROFIBUS interface absent or not operating.
E-dnEt DEVICENET interface absent or not operating.
ϩϩϩϩϩϩ Dash that runs along the perimeter of the display: BLIND function enabled.
ErrNEN Memory error. Press the PRG key to reset the memory and return the parameters to their
default values. NOTE: it is also deleted the calibration performed.