System Function Checklist
Ensure that the wall support is adequate and that the system is properly
mounted to the wall.
Ensure that all certified components bear labels that include the model
and serial number, date of manufacture and a statement of certification
as noted elsewhere in this manual.
Indication of
Technique
Factors
Verify that the “70 kV – 7 mA” is legible on the label of the Operator
Panel.
Check for oil leaks or other evidence that could indicate internal damage.
Replace the Tubehead, if necessary.
Ensure that the Tubehead maintains its position around the horizontal
axis while remaining easy to rotate and position. Also, check the vertical
pivot of the Tubehead for easy movement while remaining in position
after movement.
Check that all movements are smooth and quiet. Verify that the
Tubehead is properly counterbalanced for vertical drift and that the
Horizontal and Articulating Arms do not drift horizontally.
Verify that the switch is working properly and that the Ready Indicator is
illuminated when the power switch is in the ON position.
With the power switch, located at the base of the Control Unit, in the ON
position, verify that a time indicates on the Operator Panel. Also, check
the function of the selection buttons for Exposure Time, Tooth Selection,
Image Receptor Type and Patient Size. Pressing a selection button
should cause indicator lamps to indicate the selected item.
Verify that the Exposure button on the Operator Panel is functioning
properly. To make an exposure, press and hold the Exposure button until
the Radiation Indicator is extinguished and the audible signal is no
longer heard.
Make several exposures and verify that the Radiation Indicator
illuminates and the audible signal is heard.
Premature
Termination
Indicator
(Optional)
Select the longest exposure time possible using the Exposure Time
buttons. Initiate an exposure but release the Exposure button after a
brief period of time before the timer terminates the exposure. Verify that
the exposure time flashes, indicating a premature termination. Press the
up or down time arrows to clear the flashing time display.
Coil-cord Hand
Switch (Optional)
If a coil-cord hand switch is used, inspect the switch housing and coil
cord for damage or wear. Replace if evidence of damage is present.
Make certain that the user of the system has received a copy of the User
Manual.