RF emissions CISPR 11 Class B
The Prolife PA3 Backlight is suit-
able for use in all establishments,
including domestic estab-
lishments and those directly
connected to the public low
voltage power supply network
that supplies buildings used
for domestic purposes
Harmonic emissions
IEC 61000–3–2
Not applicable
Voltage fluctuations / flicker
IEC 61000–3–3
Not applicable
Electromagnetic Immunity: (IEC60601–1–2)
Immunity test
IEC60601–1–2
test level
Compliance level
Electromagnetic envi-
ronment –guidance
Electrostatic
discharge (ESD) IEC
61000–4–2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic
tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%
Electric fast tran-
sient / burst
IEC 61000–4–4
±2 kV for power
supply lines
±1 kV for
input / output
lines
Not applicable Mains power quality
should be that of
a typical commercial
or hospital
environment
Surge
IEC 61000–4–5
±1 kV differential
mode
±2 kV common
mode
Not applicable
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000–4–11
<5% UT
(95% dip in UT)
for 0.5 cycle
40 % UT
(60% dip in UT)
for 5 cycles
70 % UT
(30% dip inUT)
for 25 cycles
<5 %
(UT 95% dip inUT)
for 5 sec
Not applicable Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the upper
arm stlye requires
continued operation
during power mains
interruptions, it is
recommended that
the Prolife PA3 Back-
light be powered
from an uninterrupt-
ible power supply
or a battery