The Digital Blood Pressure Monitor Prolife PS1 Standard is intended for use in
the electromagnetic environment specified below. The customer or the user of
the Prolife PS1 Standard should assure that it is used in such an environment.
Electromagnetic emissions: (IEC 60601-1-2)
Emission Test Compliance Electromagnetic environment
RF emission CISPR 11 Group 1
The Prolife PS1 Standard uses RF energy
only for internal functions. Therefore, this RF
emission is extremely weak and there is little
chance of it creating any kind of interference
whatsoever with nearby electronic equipment
RF emissions CISPR 11 Class B
The Prolife PS1 Standard is suitable for use in
all establishments, including domestic estab-
lishments and those directly connected to the
public low voltage power supply network that
supplies buildings used for domestic purposes
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker IEC 61000-3-3
Not applicable
Electromagnetic immunity: (IEC 60601-1-2)
Immunity test
IEC 60601-1-2
test level
Compliance
level
Electromagnetic envi-
ronment – guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are cov-
ered with synthetic
material, the relative
humidity should be at
least 30 %
Electric fast transient/
burst IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
Not applicable Mains power quality
should be that of a typ-
ical commercial or hos-
pital environment
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
Not applicable