Loading...
10
5.3. FIXATION OF THE CUFF
Insert the connector into the cuff connection opening (as shown in Fig. 1) and
make sure that the connection is tight..
1. Fix the cuff with Velcro fastener so that it lies
comfortably and fits snugly to your arm. The
distance between the edge of the cuff and the
middle of the elbow should be about 1–2 cm.
2.
Lay the arm on a table, with the palm upwards.
The cuff should be at the level of the heart, if nec
-
essary place a cushion under the forearm. Ensure
support of the arm on a horizontal surface. Stay
calm for 2 minutes before measurement
.
3. Keep your legs straight and uncrossed, lean on
the back of the chair.
Picture 1
Connector
Cuff
Cuff connection
opening
1–2 cm
Chief artery

Table of Contents