26
TENS+EMS SUPER DUO STIMULATOR
TECHNICAL SPECIFICATIONS
Channel: Dual, isolated between channels.
Output Voltage: 0~40V peak into 500Ω load each channel.
Pulse Amplitude: Adjustable0–80mA.
Pulse Rate: As pre-programming operation mode.
Pulse Width: As pre-programming operation mode.
Software ramp up feature: Pulse width ramp up when change mode.
Timer : 5~60 min. selectable.
LCD: Show modes, pulse rate, pulse width, timer, CH1/CH2, intensity level.
Wave Form: SymmetricalBi-Phasicsquarepulse.
Max Charge per Pulse: 20.8 micro-coulombs maximum.
** All electrical speci cations are ±20%
Operating Conditions: + 50°F (10
°C) to +104° (40°C), 40-90% max. Relative humidity
Transport and Storage Temperature: +14°F (-10
°C) to +140° (60 °C), 30-95% max. Relative
humidity
Weight: 120 g (battery included)
Dim. : 122 x 66 x 30 mm
Power Source: 3 x AAA / 4.5 Volt batteries
Technical Speci cations
This symbols means “ Serial number “
This symbols means “ Attention, consult the accompanying
documents”
Thissymbolsmeans“Manufacturer“
ThissymbolmeanstypeBFequipment;thisdeviceoers
protection against electrical shock by standard compliance to
leakage currents of electrode pad.
This device shall be disposed in accordance with national laws
after their useful lives
Thissymbolmeans“usedbefore”,representas“YYYY-MM”
(for year and month).