6
7
External power supply (AC/DC) is to be connected to “Power” (10) 
jack located on the bottom side of the device.
Please note that the central pin of the power supply that you connect to 
the “power” jack of the riflescope, must have marking “+”.  The power 
supply may have marking      
EXTERNAL POWER SUPPLY
7
Connection of an external power supply (icon        appears in the status bar) automatically 
cuts off power supply from batteries.
External power supply DOES NOT charge the rechargeable batteries in the unit.
Attention! We suggest that you use battery packs EPS3 or EPS5 ensuring from 9 to 20 
hours of operation.
The scope can be powered with an external DC power supply (2.1mm 
pin) with stabilised voltage ranging from 8.4V to 15V (max. consumption 
power 3W) or a 12V vehicle socket.
OPERATION
8
Switching on and calibration
Press “ON/OFF” (1) to switch on the unit. Green LED indicator (14) will 
light up. In case of low battery the LED will turn red and      battery icon       
will start flashing.    
Do the calibration of image. Calibration levels background temperature 
of the microbolometer and eliminates image flaws.  
There are three calibration modes:
manual (M), semiautomatic (H) and automatic (А).
Manual calibration mode
The lens cap (4) should be closed (in 19mm models a diaphragm (15) 
inside the objective lens plays the role of the cap. Rotate the wheel (16) 
clockwise to open the diaphragm, counterclockwise to close it). Turn on 
the unit, press and hold the controller (5) for two seconds to enter the 
menu.
Rotate the controller to select option Cal. Press the controller.
Rotate the controller to select mode М. Press the controller to confirm. 
To exit the menu, press and hold the controller (5) for two seconds or 
wait 10 seconds for automatic exit.
Press the CAL (2) button to calibrate. The image will freeze for 1-2 
seconds. Then open the lens cap. Calibration is completed. 
In case you see image flaws (such as frozen image, vertical stripes 
etc.) re-calibrate the unit.
Semiautomatic calibration mode
Turn on the unit, open the lens cap. 
Press and hold the controller (5) for two seconds to enter the menu.
Rotate the controller to select option Cal. Press the controller.
Rotate the controller to select mode H. Press the controller to confirm. 
Press the CAL (2) button to calibrate. The image will freeze for 1-2 
seconds and you will hear the sound of the internal shutter. Calibration 
is completed.
Automatic calibration mode
With the automatic calibration mode the thermal imager calibrates by 
itself  according  to  the  software  algorythm.  The  detector 
(microbolometer)  is  closed  with  the  shutter  automatically.  User 
assisted calibration with the CAL (2) button is allowed in this mode.   
Turn on the unit, open the lens cap. 
Press and hold the controller (5) for two seconds to enter the menu.
Rotate the controller to select option Cal. Press the controller.
Rotate the controller to select mode A. Press the controller to confirm. 
At the moment of the automatic calibration the image will freeze for 1-2 
seconds and you will hear the sound of the internal shutter. 
Time  intervals  between  calibrations  depends  on  the  heat  of  the 
detector.  The  unit  may  require  calibration  more  frequently  on 
activation. 
WARNING! Do not point the objective lens of the unit at intensive 
sources  of  light  such  as  the  sun.  This  may  disable  electronic 
components  of  the  unit.  The  warranty  does  not  cover  damage 
caused by improper operation. 
Turn the lever of the battery compartment (12) 90 degrees in “Open” 
position  and,  pulling  by  the  lug  of  the  cover,  remove  the  battery 
container.  
Install four AA batteries (or rechargeable batteries) observing polarity 
shown on the battery container.  
Insert the battery container into the battery compartment and turn the 
lever 90 degrees clockwise. 
Battery charge level is displayed in the status bar (       ). 
In case of complete battery discharge, icon    is flashing in the status 
bar and in the centre of the display.
INSTALLATION OF BATTERIES
Note:  to  ensure  long  and  reliable  operation  it  is  recommended  that  you  use  quality 
rechargeable batteries with a capacity of at least 2500 mAh. Please do not use batteries of 
different types or batteries with various charge levels.
6
Focusing and image adjustment
Open the lens cap (4).
Adjust sharp image of the display icons by turning the eyepiece diopter 
adjustment ring (13). 
To adjust display brightness, press briefly the controller (5) - brightness 
icon    should appear in the upper right portion of the display. Rotate 
the controller to select brightness level (from 0 to 20).    
To adjust display brightness, press briefly the controller (5) - contrast 
icon    should appear in the upper right portion of the display. Rotate 
the controller to select contrast level (from 0 to 20).