Questions and Answers

  • M
    Michael BennettAug 20, 2025
    Why is there odour or particles in exhaust air from my Purex iFume 200, even without machine alerts?
    • T
      turnerhannahAug 20, 2025
      The exhaust monitor unit inside your Purex Air Cleaner might have a fault. Begin by inspecting all cables and connections. If the issue continues, contact your supplier.
  • A
    ashleythompsonAug 22, 2025
    What to do if Temperature (red) alert appears on Purex iFume 200 Air Cleaner?
    • C
      Carl RamirezAug 22, 2025
      If a Temperature (red) alert appears, first ensure that nothing is blocking the cooling vents located at the rear of the extractor. Then, check the process temperature and turn off the extractor.
  • J
    Juan LambertAug 25, 2025
    What to do if Temperature (yellow) alert appears on Purex iFume 200?
    • B
      Brandon RushAug 25, 2025
      If you receive a Temperature (yellow) alert, check that the cooling vents on the rear of the extractor are free from obstructions. Also, check the process temperature.
  • R
    Richard HarperAug 26, 2025
    What to do if my Purex iFume 200 screen is not illuminated?
    • D
      Daniel GilbertAug 26, 2025
      If the screen is not illuminated, there could be several reasons: * The power supply might be absent or incorrect. Check all cables and connections, ensuring the power supply matches the machine's requirements. * The circuit breaker might have tripped. Try resetting the breaker on the rear of the machine. * The control PCB inside the machine may have developed a fault. Check all cables and connections first. Then contact your supplier.
  • K
    Kathleen LopezAug 27, 2025
    What to do if Particulate (red) alert appears on Purex iFume 200?
    • P
      Peggy ChungAug 27, 2025
      If a Particulate (red) alert appears on your air cleaner, replace the main filter.
  • B
    boonedennisAug 29, 2025
    What to do if GAS (red) alert appears on Purex Air Cleaner?
    • M
      michael80Aug 29, 2025
      If a GAS (red) alert appears, you should replace the main filter.
  • B
    Brandon WoodAug 31, 2025
    Why does my Purex iFume 200 filter have a shorter than expected life?
    • G
      Gina DuarteAug 31, 2025
      A shorter than expected filter life may be due to: * A problem with the pre-filter. Check the pre-filter for damage or if it is missing. * Changes in the process or its duration, including the use of different materials. Investigate any changes in these factors. If the issue continues, contact your supplier.
  • D
    David HallSep 2, 2025
    What to do if Particulate (yellow) alert appears on Purex iFume 200?
    • D
      Douglas BurtonSep 2, 2025
      If you see a Particulate (yellow) alert, order a replacement main filter.
  • I
    ilarsonSep 4, 2025
    What to do if GAS (yellow) alert appears on Purex iFume 200 Air Cleaner?
    • G
      Gary GlennSep 4, 2025
      If a GAS (yellow) alert appears, order a replacement main filter.
  • T
    Todd LeeSep 6, 2025
    What to do if the Purex iFume 200 control signal to/from interfaced equipment is incorrect?
    • K
      Kenneth MontgomerySep 6, 2025
      If the control signal to or from interfaced equipment is incorrect, and the equipment isn't switching on and there's an alarm, it indicates an electrical problem. Check all cables, connections, and the interfaced equipment.