5 Instrument Overview
5.1 Control Keys
The SP-350 has three control keys, as shown in Figure 2. The le ( ), right ( ), and ok ( ) keys
are used to launch acons indicated on the LCD display directly above the keys. The labels above the keys
indicate the funcon associated with each key and funcons can be changed in dierent operaon modes.
Figure 2.
5.2 Main Module On/O
To turn on the SP-350: Press momentarily and release.
To turn o the SP-350: Press and hold for about three seconds. Release when the LCD display
turns o. The SP-350 turns itself o aer 60 seconds without user interacon detected. This is done to
conserve baery life.
*
NOTE
*
This auto-me o seng may be customized by the user as desired through the
uPyxis® Mobile or Desktop App.
6 Measurement
6.1 PTSA Measurement
When powered on, the SP-350 will default to the PTSA measurement mode. The water sample can be
transferred to the sample cell using a pipee or lled directly from a faucet, sample bole, or sample
valve.
*
NOTE
*
Special care should be taken when pouring the sample into the cell to avoid air
bubble entrainment, which can interfere with reading accuracy.
Before beginning a measurement, use the sample water to rinse the sample cell at least three mes. Allow
5–10 seconds for the SP-350 to stabilize. The values will be displayed in white with a blue background if
a stable value is reached (Figure 2). For a sample containing 100 ppb PTSA, the measured PTSA should be
stabilized within the range of 98–102 ppb.
SP-350 User Manual service@pyxis-lab.com | +1 (866) 203-8397 5