NOTE
*
The me required to reach a stable reading may be slightly longer if the water
sample temperature is signicantly dierent than the environmental temperature at which
the SP-350 had been equilibrated (stored).
6.2 High Color and Turbidity Warning
The SP-350 has extra channels to measure sample turbidity and color to automacally compensate sam-
ple color and turbidity interference. If sample turbidity and color values determined are too high, a PTSA
measurement warning will be displayed. In such a case, the user should lter the sample for PTSA mea-
surement.
7 Calibraon
7.1 PTSA Calibraon (Two-Point with Zero)
1. Rinse the sample cell three mes with DI water. Fill the sample cell with DI water.
*
NOTE
*
In emergency, “non-PTSA” water, such as city water, may be used, but re-
calibrate using DI water for the zero step as soon as it is available.
2. Power on the SP-350 by pressing . Allow 5–10 seconds for the SP-350 to stabilize.
3. The unit is acvely reading and displaying PTSA. The values will be very low if DI water is used; PTSA
value should be near zero. A low non-zero value (e.g. 0.2 or 0.4, etc.) is not problemac.
4. Press P-Cal ( or ) to launch the PTSA CALIBRATION screen (Figure 3).
5. Press Zero ( ) to start the zero (blank) calibraon.
6. If the calibraon succeeds, a checkmark ( ) and instrucons for the slope calibraon will appear
(Figure 4).
7. Rinse the sample cell three mes with the desired PTSA standard. Fill the sample cell with the desired
PTSA standard.
8. Press Cycle ( ) to cycle between the PTSA standards 100, 200, and 300 ppb (it repeats). Ensure
the value selected matches the desired PTSA standard in the sample cell.
9. Press Slope ( ) to start the slope calibraon.
10. If the calibraon succeeds, a checkmark ( ), a ”Calibraon Success” message will appear (Figure
5). Otherwise, a warning message is displayed.
11. Calibraon is now complete. Press Exit ( ) to return to measurement mode.
*
NOTE
*
If Exit is pressed before the second checkmark appears, the calibraon will
not be completed and must be re-done.
SP-350 User Manual service@pyxis-lab.com | +1 (866) 203-8397 6