Page 19 of
Pulse on 10 sec
Pulse off __ sec
Pulse on 10 sec
Pulse off 10 sec
Time 0:30:00
Pulse 10 10 Ampl 40 %
2. Using the numeric keys, set the ON portion of the cycle, then press the ENTER/REVIEW key.
The screen will display:
3. Using the numeric keys set the OFF portion of the cycle, then press the ENTER/REVIEW key.
The screen will display:
4. Press the ENTER/REVIEW key.
The screen will display:
On Cycle Off Cycle
REVIEW: The REVIEW function provides a “window” on the process by displaying various
operating parameters without process interruption. Pressing the ENTER/REVIEW key repeatedly
during processing will consecutively display the following information.
a) Selected amplitude:
e.g. Amplitude 40%
b) Selected processing time and elapsed processing time:
e.g. Set 0:30:00 Time 0:15:00
c) Selected pulsing cycle and actual pulsing cycle:
e.g. Pulse 10 10 / (actual time)
d) Amount of power in watts, and accumulated amount of energy in JOULES delivered to
the probe (Note: The amount of energy displayed will be only for one cycle. Initiating
a new cycle will reset the display to zero.):
e.g. 20 watts / XXX Joules
e) Elapsed time since processing was initiated:
e.g. Elapsed time 0:15:00