9.2. ATTACHING & DETACHING A PROBE
1. Disconnect probe from converter. Use the wrench set provided with the system.
2. Clean threaded stud. Use alcohol and a cotton swab to remove any debris on the threading
of the connecting stud. Allow the alcohol to dry completely.
3. Clean threading in converter. Hold the converter horizontally and use alcohol and a cotton
swab to remove any debris on the threading. Do not allow liquid to drip into Converter.
Allow the alcohol to dry completely.
4. Reattach probe to converter. Screw the probe back onto the converter and tighten with the
wrench set provided.
5. The tips on replaceable tip probes can be removed for cleaning and/or replacement. When
replacing horns or horn tips, always clean the threaded mating surfaces of the converter
and horn. Use alcohol and a cotton swab to remove any debris on the threading of the tip or
probe. If the replaceable tip loosens during sonication, be sure to remove the tip for
cleaning and inspect the threading on the tip and probe. Call the manufacturer for
assistance if the threading is chipped or damaged in anyway.