Page 30 of
12.2 COOLING AIR REGULATION AND ADJUSTMENT
Adjust the regulated compressed air until the gauge indicates 10 ± 1 PSIG (approximately 4
CFM). The compressed air will flow into the converter and out through the outlet barb. Verify the
compressed air is flowing out from the exhaust tube upon each use.
Example of the air cooling hoses set up
properly with the enclosure.