Replacement Tip Removal Replacement Tip Tightening
Microtip Removal Microtip Tightening
Note: When tightening a Microtip the tip must not be in contact with the work surface. Always
have the tip extending off of the table or work surface to minimize stress to the tip.
Note: For additional information on tightening and removing probes, please visit the Videos
section of the Qsonica website. https://www.sonicator.com/pages/videos
9.3. SYSTEM CLEANING INSTRUCTIONS
The power supply and converter may be cleaned using an acid-free cleaning solution (i.e. glass
cleaner).
Probes should be cleaned using isopropyl alcohol. Probes are made from titanium and can be
autoclaved (the converter is an electrical part and cannot be sterilized in this manner). Before
each procedure place the probe tip in water or alcohol and turn the power on for a few seconds to
remove residue. The tip also can be sterilized using alcohol with the power on.