MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
26, 27
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
28, 29
System Config mismatch Error
Call Service
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
30
Fluid Level Sensor test failure
Manually Vent bubble trap
Proceed with Caution
RESTART CONFIRM
* Continue with case while
manually venting air
accumulations in bubble trap
* Contact service following
procedure
31
Air-In-Line Sensor test failure
Air detection may be compromised
Proceed with caution
RESTART CONFIRM
* Continue with case while
manually monitoring for air in
delivery line
* Contact service immediately
following procedure
32
Is there fluid in
the bubble trap?
RETEST YES
* If fluid is present, select YES
OR
* If fluid is not present, restart
and
repeat the priming process &
contact service after multiple
occurrences
33
Is there fluid in
the delivery line?
RETEST YES
* If fluid is present, select YES
OR
* If fluid is not present, restart
and
repeat the priming process &
contact service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action
7

Other manuals for Quest Engineering MPS 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quest Engineering MPS 3 and is the answer not in the manual?

Quest Engineering MPS 3 Specifications

General IconGeneral
BrandQuest Engineering
ModelMPS 3
CategoryMedical Equipment
LanguageEnglish