MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
34
Air in bubble trap after Prime
Check source fluid and circuit
RETEST
* If there is NOT air present in
heat exchanger, check heat
exchanger to ensure it is
secured tightly with the
locking
knob & Reprime the circuit
OR
* If there is air present in heat
exchanger, check the lines for
clamps and ensure there is
adequate blood source
pressure & Reprime the
circuit
35
Delivery Set pressure test error
Check circuit and
Delivery/Vent valves
RETEST
* Check heat exchanger to
ensure it is secured tightly
with
the locking knob
* Manually actuate the delivery
and vent valves & Reprime
the
circuit
* Check lines for clamps, kinks
and occlusions
36
Heat Exchanger pressure test failed
Replace delivery set
REPRIME
* Check heat exchanger to
ensure it is tightly secured
with
the locking knob
* If using a heater/cooler unit,
turn off heater/cooler unit and
repeat the prime sequence
* Replace disposable set after
multiple occurrences
37
Unable to Prime due to inadequate fill
Check fluid source pressure and circuit
RETEST
* Check the lines for clamps,
ensure there is adequate
blood
source pressure & Reprime
the
circuit
38
Unable to Prime
Check fluid source pressure and circuit
RETEST
* Check the lines for clamps,
ensure there is adequate
blood
source pressure & Reprime
the
circuit
Alar
m
Cod
e
Message Display
Corrective Action
8