MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
Delivery occlusion
Check for clamps
Check pressure limit setting
CONFIRM
* Check delivery line for
clamps,
kinks, or occlusions
* Check the upper pressure
limit
settings
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
Inadequate Crystalloid fill
Check Source
CONFIRM
* Check Crystalloid inlet line for
clamps, kinks, or occlusions
* Raise the Crystalloid bag or
use
a pressure cuff
Inadequate Blood fill
Check Source
CONFIRM
* Check Blood inlet line for
clamps, kinks, or occlusions
* Increase the blood inlet
source
pressure
Inadequate Fill Check Source
CONFIRM
* Check blood/crystalloid
source
lines and pressures
* If problem persists, restart
system
* Call service after multiple
occurrences
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
System FPGA Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
System FPGA Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
System Mechanical Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences