MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
Door is open
Close door to continue
SHUTDOWN
* Close the door
* Press shutdown to turn
System
off
System FPGA Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
Blood Pump Error
Flow may be resumed
SHUTDOWN CONFIRM
* Continue with procedure
* If error occurs multiple times
over the course of the
procedure, contact service
immediately following
procedure
Unable to maintain flow rate setting
Check circuit
CONFIRM
* Check blood/crystalloid
source
lines and pressures
* Stop flow and restart flow
* If problem persists, restart
system
* Call service after multiple
occurrences
Mechanism valve error
Reinitializing. Please wait…
* Continue with procedure
* If error occurs multiple times
over the course of the
procedure, contact service
immediately following
procedure
Delivery valve sensor mismatch
Flow may be resumed
CONFIRM
* Continue with procedure
* If error occurs multiple times
over the course of the
procedure, contact service
immediately following
procedure
Arrest pump failed diagnostics
Proceed without Arrest pump
RESTART CONFIRM
* Restart System
* Call service after multiple
occurrences