MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
99
Arrest cartridge is empty
Fill Arrest cartridge, press RETEST
DISABLE RETEST
* Choose DISABLE if Arrest
Cartridge is no longer needed
for procedure
OR
* Refill Arrest Cartridge and
select RETEST
100
Arrest cartridge not detected
Install Arrest cartridge, press RETEST
DISABLE RETEST
* Check Arrest Cartridge to
verify
it is properly installed and fully
secured to the pump
* If the cartridge was secure, try
removing cartridge and
reinserting it
* If error still persists, restart
console and try again
* Contact service after multiple
attempts
101
Arrest occlusion during prime
Check Arrest line and stopcock
CONFIRM RETEST
* Check arrest delivery line for
clamps, kinks, or occlusions
* Check stopcock position
* If problem persists, replace
arrest delivery line and try
again
* If problem still exists after
replacing arrest delivery line,
restart system and try again
* Contact service after multiple
occurrences
102
Arrest cartridge is empty
Arrest pump is Off
REFILL CONFIRM
* Choose CONFIRM if Arrest
Cartridge is no longer needed
for procedure
OR
* Select REFILL and refill Arrest
Cartridge
Alar
m
Cod
e
Message Display
Corrective Action