MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
103
Arrest occlusion detected
Check Arrest line and stopcock
Arrest pump is Off
CONFIRM RETEST
* Check arrest delivery line for
clamps, kinks, or occlusions
* Check stopcock position
* If problem persists, replace
arrest delivery line and try
again
* If problem still exists after
replacing arrest delivery line,
restart system and try again
* Contact service after multiple
occurrences
104
Arrest over delivery detected
Arrest has been turned Off
RETEST CONFIRM
* Select CONFIRM to turn
Arrest
pump off and continue
procedure
OR
* Select RETEST to turn Arrest
pump on and continue
procedure
105
Arrest under delivery detected
Check cartridge and circuit
CONFIRM
* Check arrest delivery line for
clamps, kinks, or occlusions
* Check cartridge for
occlusions/leaks
* Check stopcock position
* If problem persists, replace
arrest delivery line and try
again
* If problem still exists after
replacing arrest delivery line,
restart system and try again
* Contact service after multiple
occurrences
106, 107,
108
Additive pump failed diagnostics
Proceed without Additive pump
RESTART CONFIRM
* Restart System
* Call service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action
16