MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
Additive cartridge is empty
Fill Additive cartridge, press RETEST
DISABLE RETEST
* Choose DISABLE if Additive
Cartridge is no longer needed
for procedure
OR
* Refill Additive Cartridge and
select RETEST
Additive cartridge not detected
Install Additive cartridge, press RETEST
DISABLE RETEST
* Check Additive Cartridge to
verify it is properly installed
and
fully secured to the pump
* If the cartridge was secure, try
removing cartridge and
reinserting it
* If error still persists, restart
console and try again
* Contact service after multiple
attempts
Additive occlusion during prime
Check Additive line and stopcock
CONFIRM RETEST
* Check additive delivery line
for
clamps, kinks, or occlusions
* Check stopcock position
* If problem persists, replace
arrest delivery line and try
again
* If problem still exists after
replacing additive delivery
line,
restart system and try again
* Contact service after multiple
occurrences
Additive cartridge is empty
Additive pump is Off
REFILL CONFIRM
* Choose CONFIRM if Additive
Cartridge is no longer needed
for procedure
OR
* Select REFILL and refill
Additive Cartridge