MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
Arrest cartridge not detected
Verify Arrest cartridge installation
Arrest pump is Off
CONFIRM RETEST
* Check Arrest Cartridge to
verify
it is properly installed and fully
secured to the pump
* If the cartridge was secure, try
removing cartridge and
reinserting it
* If the cartridge was secure, try
stopping and restarting
cardioplegia flow
* If error still persists, restart
console and try again
* Contact service after multiple
occurrences
Arrest pump piston sensor error
Verify Arrest cartridge installation
CONFIRM RETEST
* Check Arrest Cartridge to
verify
it is properly installed and fully
secured to the pump
* If the cartridge was secure, try
removing cartridge and
reinserting it
* If the cartridge was secure, try
stopping and restarting
cardioplegia flow
* If error still persists, restart
console and try again
* Contact service after multiple
occurrences