Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
966User Manual 1178.8520.02 ─ 01
CALCulate<n>:SGRam[:STATe].......................................................................................968
CALCulate<n>:SPECtrogram[:STATe].............................................................................. 968
CALCulate<n>:SGRam:THReedim[:STATe]...................................................................... 968
CALCulate<n>:SPECtrogram:THReedim[:STATe]..............................................................968
CALCulate<n>:SGRam:TRACe....................................................................................... 969
CALCulate<n>:SPECtrogram:TRACe...............................................................................969
CALCulate<n>:SGRam:TSTamp:DATA?...........................................................................969
CALCulate<n>:SPECtrogram:TSTamp:DATA?.................................................................. 969
CALCulate<n>:SGRam:TSTamp[:STATe]..........................................................................970
CALCulate<n>:SPECtrogram:TSTamp[:STATe]................................................................. 970
CALCulate<n>:SGRam:CLEar[:IMMediate]
CALCulate<n>:SPECtrogram:CLEar[:IMMediate]
This command resets the spectrogram and clears the history buffer.
Example:
CALC:SGR:CLE
Resets the result display and clears the memory.
Manual operation: See " Clear Spectrogram " on page 389
CALCulate<n>:SGRam:CONTinuous <State>
CALCulate<n>:SPECtrogram:CONTinuous <State>
This command determines whether the results of the last measurement are deleted
before starting a new measurement in single sweep mode.
This setting applies to all spectrograms in the channel.
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
INIT:CONT OFF
Selects single sweep mode.
INIT;*WAI
Starts the sweep and waits for the end of the sweep.
CALC:SGR:CONT ON
Repeats the single sweep measurement without deleting the
results of the last measurement.
Manual operation: See " Single Sweep / Run Single " on page 387
See " Continue Frame " on page 389