Functionality Outside the Secured Area
R&S
®
FSVA3000 / R&S
®
FSV3000
10Instrument Security Procedures 1179.0097.02 ─ 01
6 Functionality Outside the Secured Area
To establish the functionality outside the secured area:
1. Configuration without option R&S FSV3-B20 (removable hard drive)
a) Turn on the R&S FSV/A.
The internal SSD memory module holds all instrument data for operation.
2. Configuration with option R&S FSV3-B20 (removable hard drive)
As the CFAST memory card holds the operating system, the R&S FSV/A cannot be
operated without a CFAST memory card. For servicing and calibration,
Rohde & Schwarz provides a separate removable CFAST memory card (option
R&S FSV3-B18). This CFAST memory card contains the operating system and
required instrument data.
To establish the functionality:
a) Insert a second CFAST memory card (R&S FSV3-B18).
This memory card enables the R&S FSV/A to start the operating system.
b) Turn on the R&S FSV/A.
The instrument is ready for use.
7 Validity of Instrument Calibration after
Declassification
Calibration makes sure that measurements comply to the specified characteristics.
Rohde & Schwarz recommends that you follow the calibration cycle suggested for your
instrument.
The R&S FSV/A uses two different sets of alignment data.
Factory alignment data
These data are saved on EEPROMs. Therefore, replacing one removable CFAST
memory card with another card does not affect the availability of the factory alignment
data.
User action: none required.
Self alignment data
These data are saved on the CFAST memory card or the SSD.
User action: In the following cases, user action is required:
●
When the CFAST memory card was replaced, the self alignment has to be run
once.