Instrument Declassification
R&S
®
FSVA3000 / R&S
®
FSV3000
8Instrument Security Procedures 1179.0097.02 ─ 01
– If R&S FSV3-K33 is available with Secure User Mode enabled:
The R&S FSV/A redirects user data and instrument setups to the volatile mem-
ory (SDRAM).
Sanitization procedure: Turn of instrument power.
●
If the CFAST memory card is not inserted:
Without the CFAST memory card, holding the operating system etc., you cannot
work with the R&S FSV/A.
Sanitization procedure: None required.
See also Chapter 5, "Instrument Declassification", on page 8.
5 Instrument Declassification
Before you can remove the R&S FSV/A from a secured area (for example to perform
service or calibration), all classified user data needs to be removed.
Overview for declassification of the R&S FSV/A depending on the configuration
The options R&S FSV3-B20 (removable hard drive) and R&S FSV3-K33 (security write
protection of solid state drive) determine the sanitization procedures for the
R&S FSV/A as shown in Figure 5-1.
Refer also to Types of Memory and Information Storage in the R&S FSV/A, Secure
User Mode, and chapter Protecting Data Using the Secure User Mode in the user man-
ual of the R&S FSV/A, see www.rohde-schwarz.com/manual/fsv3000/.
Figure 5-1 illustrates the different ways of data storage and sanitization procedures,
depending on the options the instrument is equipped with.