Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1045User Manual 1178.8520.02 ─ 01
<li> Limit line
Parameters:
<Mode> ABSolute
Limit line is defined by absolute physical values (Hz or s).
RELative
Limit line is defined by relative values related to the center fre-
quency (frequency domain) or the left diagram border (time
domain).
*RST: ABSolute
CALCulate<n>:LIMit<li>:CONTrol:OFFSet
<Offset>
This command defines an offset for a complete limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Offset> Numeric value.
The unit depends on the scale of the x-axis.
*RST: 0
Default unit: HZ
Manual operation: See " X-Offset " on page 476
CALCulate<n>:LIMit<li>:CONTrol:SHIFt <Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Distance> Numeric value.
The unit depends on the scale of the x-axis.
Default unit: HZ
Manual operation: See " Shift x " on page 480
CALCulate<n>:LIMit<li>:CONTrol:SPACing <InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from one horizontal point to the next.