EasyManua.ls Logo

R&S RTO - Page 173

R&S RTO
829 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurements
R&S
®
RTO
157User Manual 1316.0827.02 ─ 06
For a description of "Selection" settings, see "Selection Settings" on page 153.
Hysteresis
Defines a hysteresis for the middle reference level. A rise or fall from the middle reference
value that does not exceed the hysteresis is rejected as noise.
SCPI command:
REFLevel<m>:RELative:HYSTeresis on page 537
Tube Tab
This tab allows you to define evaluation tubes for measurements that require high level
or low level detection. If the signal value remains within the defined tubes, it is considered
to be high or low.
For a description of "Selection" settings, see "Selection Settings" on page 153.
Top outer
Defines an area above the high signal level which is still considered to be high level.
SCPI command:
REFLevel<m>:ABSolute:TOTube on page 538
MEASurement<m>:REFLevel:RESult:TOUTer on page 541
Top inner
Defines an area beneath the high signal level which is still considered to be high level.
SCPI command:
REFLevel<m>:ABSolute:TITube on page 539
MEASurement<m>:REFLevel:RESult:TINNer on page 541
Automatic Measurements

Table of Contents

Related product manuals