Remote Control
R&S
®
RTO
559User Manual 1316.0827.02 ─ 06
Parameters:
<Mode> ABS | REL
*RST: ABS
Parameters for setting and query:
<HistogramName>
LAYout:HISTogram:VERTical:ABSolute:STARt <HistogramName>,<Start>
LAYout:HISTogram:VERTical:ABSolute:STARt? <HistogramName>
Defines the vertical start value of the histogram.
Parameters:
<Start> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0.01
Parameters for setting and query:
<HistogramName>
LAYout:HISTogram:VERTical:ABSolute:STOP <HistogramName>,<Stop>
LAYout:HISTogram:VERTical:ABSolute:STOP? <HistogramName>
Defines the vertical stop value of the histogram.
Parameters:
<Stop> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0.01
Parameters for setting and query:
<HistogramName>
LAYout:HISTogram:VERTical:RELative:STARt <HistogramName>,<RelativeStart>
LAYout:HISTogram:VERTical:RELative:STARt? <HistogramName>
Defines the vertical start value of the histogram.
Parameters:
<RelativeStart> Range: -100E+24 to 100E+24
Increment: 0.1
*RST: 0
Default unit: %
Parameters for setting and query:
<HistogramName>
Usage: SCPI confirmed
Command Reference