Remote Control
R&S
®
RTO
584User Manual 1316.0827.02 ─ 06
The higher the overlap factor, the more frames are used. This leads to more individual
results and improves detection of transient signal effects. However, it also extends the
duration of the calculation.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<OverlapFactor> Range: 0 to 90
Increment: 1
*RST: 50
Default unit: %
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:GATE:ABSolute:STARt <Start>
Defines the starting value for the gate.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<Start> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0.01
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:GATE:ABSolute:STOP <Stop>
Defines the end value for the gate.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<Stop> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0.01
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:GATE:MODE <Mode>
Defines whether the gate settings are configured using absolute or relative values.
Suffix:
<m>
.
1...4
math waveform
Command Reference