Operating the instrument
R&S
®
RTO6
55Getting Started 1801.6670.02 ─ 03
One-click actions Interactive actions
Force Trigger Zone trigger
Demo Spectrogram (option R&S RTO6-K37)
You can configure the content of the toolbar, see Chapter 6.6.2, "Configuring
the toolbar", on page 53.
The following list describes at first the default toolbar functions and then the addi-
tional functions.
Undo
Undoes the last setting actions step by step. Some actions cannot be
revoked: locking the touchscreen with [Touch Lock], and saving data. The undo
stack is deleted during the following actions: Reloading settings from file, and ref-
erence waveform actions (save, load and preset with active reference waveform).
Redo
Recovers the undo steps in reverse order.
Help
Enables the tooltip display. A short description appears when you tap a
parameter in a dialog. To open the corresponding help topic, tap the "Show Help"
button in the lower right corner of the tooltip. See also: Chapter 6.11, "Getting
information and help", on page 68.
Graphical Saveset
Opens a window to select and load instrument settings that were previ-
ously stored in a saveset. A graphical preview helps you to find the required set-
tings.
Saveset
Saves the current instrument settings in a saveset.
You can reload the saveset using the "Graphical Saveset" toolbar icon, or using
"Menu" >"Save/Recall" > "Recall" tab > "Saveset with preview".
The filename is created according to the autonaming pattern, defined in "Menu" >
"Settings" > "Save/Recall" > "Autonaming" tab.