Command reference
R&S
®
ZNA
1278User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DELete:ALL
Deletes all port assignments (see Chapter 4.5.5.6, "Multiple port assignments",
on page 189).
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine
Usage: Event
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
Starts the automatic calibration sweep for the indicated channel and port assignment.
A complete, valid set of port assignments must be defined before you can initiate a cal-
ibration. See Chapter 4.5.5.6, "Multiple port assignments", on page 189 for background
information.
Suffix:
<Ch>
.
Number of the calibrated channel
<Asg> Number of the port assignment
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine
Usage: Event
Manual operation: See "Start Cal Sweep / Abort Sweep" on page 525
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
<TestPort1>, <CalUnitPort1>[, <TestPort2>]...
Manually defines port assignment no. <Asg> for channel no. <Ch>.
With manual configuration a non-minimal set of port assignments can be created,
which may increase measurement accuracy. On the other hand, it is up to the user to
ensure that the created set of port assignments is complete and valid for the chosen
calibration type (see [SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure).
Note that during the corresponding calibration sweep the R&S ZNA expects the physi-
cal port connections to be established exactly as defined by the port pairs.
Use
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:
TPORt command to take advantage of auto-detection of port connections.
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
or [SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:
TPORt:DEFault to create the default assignments instead.
SCPI command reference