Command reference
R&S
®
ZNA
1282User Manual 1178.6462.02 ─ 20
Setting parameters:
<CalName>
Example:
SENSe1:CORRection:COLLect:AUTO:CONFigure
FNPort,''
Sets the automatic calibration to "Full n-port" with factory char-
acterization.
Example: See Chapter 8.2.5.6, "Vector mixer calibration using calibration
units", on page 1622
Manual operation: See "Calibration Type" on page 530
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
<PortNumber>
Defines the "Common Port" of an automatic calibration.
This port will be part of every port assignment that is defined using [SENSe<Ch>:
]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault, or [SENSe<Ch>:
]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
on page 1277.
If [SENSe:]CORRection:COLLect:AUTO:MCONnect is ON, this port will be used as
the center of all Through measurements. If you set up the port assignments manually,
make sure that every port assignment contains it.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<PortNumber> Number of the common port
Usage: Setting only
Manual operation: See "Common Port" on page 530
[SENSe<Ch>:]CORRection:COLLect:AUTO:MIXer <Port1>, <Port2>
Acquire calibration data for the calibration mixer in an automatic vector mixer calibra-
tion
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Port1> VNA port connected to the RF port of the calibration mixer
<Port2> VNA port connected to the IF port of the calibration mixer
Example: See Chapter 8.2.5.6, "Vector mixer calibration using calibration
units", on page 1622
Usage: Setting only
Options: R&S ZNA-K5
SCPI command reference