GUI reference
R&S
®
ZNA
530User Manual 1178.6462.02 ─ 20
Common Port ← Cal. Settings
This column is only available if "Use Reduced Number of Through" on page 789 is
activated in the "Calibration" tab of the "System Config" dialog.
For full n-port automatic calibrations, it defines the port that must appear in each port
assignment, and that are as the "center" of all Through measurements.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
Calibration Type ← Cal. Settings
Displays the calibration type.
Tap the table cell to open the "Define Calibration" dialog that lets you choose calibrated
ports and type of the calibration definition. The available calibration types depend on
the channel mode of the active channel.
Figure 5-24: Define Calibration dialog: manual calibration
Cal softtool