GUI reference
R&S
®
ZNA
531User Manual 1178.6462.02 ─ 20
Figure 5-25: Define Calibration dialog: automatic calibration
The green arrow symbols give a preview of the type and the number of calibration
sweeps involved:
●
Curved arrows (example: "Refl Norm Open") denote one or more reflection mea-
surements at each port.
Reflection calibration types can be used for any set of test ports: reflection calibra-
tions are repeated for each port.
●
Straight, horizontal arrows (example: "Trans Norm") denote one or more transmis-
sion measurements between each pair of two ports.
Transmission calibration types require at least two physical ports. For unidirectional
transmission calibration types ("Trans Norm", "One Path Two Ports"), the
("Source" port) must be specified in addition.
●
The full n-port calibration types (n > 1, e.g. "TOSM") are symbolized by a closed
square symbol. The number of arrows increases the complexity but can also
improve the accuracy of the calibration.
For an overview, refer to Table 4-7.
Remote command:
Manual calibration
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
Automatic calibration
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
Power Meter ← Cal. Settings
Shows a list of all power meters that have been properly configured and are available
for the scalar source power calibration of a single source port. The last configured
power meter is selected by default. See "Configured Devices" on page 819.
Remote command:
[SENSe:]CORRection:COLLect:PMETer:ID