Command reference
R&S
®
ZNA
1296User Manual 1178.6462.02 ─ 20
Example:
*RST; SENSe1:CORR:COLL:FIXT:LMP OFF
Activate "Direct Compensation".
SENSe1:CORR:COLL:FIXT:EXP 'Traces\p12_short.csv', SHOR, 1, 2
Acquire "Direct Compensation" data for ports 1 and 2 (whose
inner connectors have to be terminated with a Short) and save
the acquired data to file.
In a subsequent measurement session you can load and apply
these data as follows:
*RST;
:SENSe1:CORR:COLL:FIXT:LMP OFF
:SENSe1:CORR:COLL:FIXT:IMP 'Traces\p12short.csv', SHOR, 1, 2
:SENSe1:CORR:COLL:FIXT:SAVE
Usage:
Setting only
Manual operation: See "Measure Fixture wizard" on page 665
[SENSe<Ch>:]CORRection:COLLect:FIXTure:SAVE
Completes a fixture compensation, storing and applying the acquired data.
Suffix:
<Ch>
.
Channel number of the corrected channel
Example: See [SENSe<Ch>:]CORRection:COLLect:FIXTure[:
ACQuire]
Usage: Event
Manual operation: See "Measure Fixture wizard" on page 665
[SENSe<Ch>:]CORRection:COLLect:FIXTure:STARt
Prepares the analyzer for fixture compensation comprising a single or a series of fix-
ture compensation sweeps
([SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]). Previous compen-
sation data is deleted.
Suffix:
<Ch>
.
Channel number of the corrected channel
Example: See [SENSe<Ch>:]CORRection:COLLect:FIXTure[:
ACQuire]
Usage: Event
Manual operation: See "Measure Fixture wizard" on page 665
SCPI command reference