GUI reference
R&S
®
ZNA
665User Manual 1178.6462.02 ─ 20
Access:
●
Channel – [Offset Embed] > "Offset" > "Fixture Compensation..."
●
Channel – [Offset Embed] > "One Way Loss" > "Fixture Compensation..."
Background Information
Refer to Chapter 4.6.1.5, "Fixture Compensation", on page 205.
Ports
Selects the ports for whom fixture compensation data shall be acquired.
Auto Length / Auto Length and Loss
"Auto Length" or "Auto Length and Loss" implies that a global electrical length offset
and loss is determined in analogy to the general offset compensation (see Chap-
ter 4.6.1.3, "Auto Length", on page 202 and Chapter 4.6.1.4, "Auto Length and Loss",
on page 204).
Remote command:
[SENSe:]CORRection:COLLect:FIXTure:LMParameter:LOSS[:STATe]
Direct Compensation
With "Direct Compensation", a frequency-dependent transmission factor is calculated;
see "Auto Length and Loss vs. Direct Compensation" on page 205.
Remote command:
[SENSe:]CORRection:COLLect:FIXTure:LMParameter[:STATe]
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:DFComp[:STATe]?
Prompt for Each Port
Determines how the R&S ZNA performs the sweeps for a given termination type (Open
and/or Short; see "Measurement Type" on page 665).
●
If unchecked, it performs the sweeps for Open/Short without interruption, implicitly
assuming that all ports are terminated accordingly
●
If checked, it interrupts the data acquisition process after each port, which allows
you to modify the test setup (e.g. terminate the next measured port).
Measurement Type
The "Open" , "Short", and "Open and Short" buttons bring up the "Measure Fixture wiz-
ard" dialog that guides you through the actual fixture measurement. See "Open/Short
vs. Open and Short compensation" on page 206.
Measure Fixture wizard
The "Measure Fixture" dialog guides you through the previously configured fixture
compensation measurements.
Offset Embed softtool