Command reference
R&S
®
ZNA
1271User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]CORRection:COLLect:DETector <CalName>[, <CalDetectorType>]
Selects the detector type during manual calibration <CalName>.
Suffix:
<Ch>
.
Channel number
Parameters:
<CalName> Name of a manual calibration previously created e.g. using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
<CalDetectorType> NORMal | AVERage
AVERage
Average detector, recommended for channels measuring traces
using detectors other than the NORMal one
Automatically selected for noise figure calibrations (see
[SENSe<Ch>:]CORRection:COLLect:NFIGure:STARt).
NORMal
Normal (sample) detector, recommended for all other applica-
tions
*RST: NORMal
Manual operation: See "Detector" on page 590
[SENSe:]CORRection:COLLect:FIXTure:LMParameter[:STATe] <Boolean>
Selects an Auto Length (and Loss) calculation or a Direct Compensation.
Parameters:
<Boolean> 1 - Auto Length (and Loss), depending on the last
[SENSe<Ch>:]CORRection:COLLect:FIXTure:
LMParameter:LOSS[:STATe] setting
0 - Direct Compensation
*RST: 1
Example: See [SENSe<Ch>:]CORRection:COLLect:FIXTure[:
ACQuire]
Manual operation: See "Direct Compensation" on page 665
[SENSe:]CORRection:COLLect:FIXTure:LMParameter:LOSS[:STATe] <Boolean>
Selects an Auto Length or an Auto Length and Loss calculation.
Parameters:
<Boolean> 1 - Auto Length and Loss
0 - Auto Length, no loss
*RST: 1
Example: See [SENSe<Ch>:]CORRection:COLLect:FIXTure[:
ACQuire]
SCPI command reference