[SENSe<Ch>:]CORRection:COLLect:NFIGure:STARt <arg0>, <arg1>, <arg2>,
<arg3>, <arg4>
Initiates a noise figure calibration for the related channel.
The calibration sweeps are started using [SENSe<Ch>:]CORRection:COLLect:
NFIGure[:ACQuire], the noise figure calibration is terminated using [SENSe<Ch>:
]CORRection:COLLect:NFIGure:SAVE or [SENSe<Ch>:]CORRection:
COLLect:NFIGure:END.
Suffix:
<Ch>
.
Number of the channel to be noise figure calibrated.
The subsequent noise figure calibration commands must use the
same channel number.
Setting parameters:
<arg0> Number of the physical VNA port connected to the DUT output.
<arg1> Number of the physical VNA port connected to the DUT input.
<arg2> ON | OFF - Boolean parameter, indicating whether an external
attenuator is used.
<arg3> Attenuation of the source level during the source noise calibra-
tion.
<arg4> Attenuation of the source level during the measurement (i.e.
after calibration).
Example: See Chapter 8.2, "Condensed programming examples",
on page 1599.
Usage: Setting only
Options: R&S ZNA-K30
[SENSe<Ch>:]CORRection:COLLect:NFIGure[:ACQuire] <arg0>
Starts one of the three noise calibration sweeps that are required for a noise figure cali-
bration of the related channel.
The noise figure calibration is initiated using [SENSe<Ch>:]CORRection:COLLect:
NFIGure:STARt and must be terminated by one of the commands [SENSe<Ch>:
]CORRection:COLLect:NFIGure:END or [SENSe<Ch>:]CORRection:
COLLect:NFIGure:SAVE.
Note: A noise figure calibration must be terminated with one of the following com-
mands, to be sent immediately after the sequence of ...[:ACQuire] commands:
●
Abort the noise figure calibration, discarding the acquired calibration data.
●
Terminate the noise figure calibration, applying the acquired calibration data (the
active channel is calibrated) These commands ensure that the analyzer is able to
start a new measurement sweep.