[SENSe<Ch>:]CORRection:COLLect:NFIGure:STARt <arg0>, <arg1>, <arg2>,
<arg3>, <arg4>
Initiates a noise figure calibration for the related channel.
The calibration sweeps are started using [SENSe<Ch>:]CORRection:COLLect:
NFIGure[:ACQuire], the noise figure calibration is terminated using [SENSe<Ch>:
]CORRection:COLLect:NFIGure:SAVE or [SENSe<Ch>:]CORRection:
COLLect:NFIGure:END.
Suffix:
<Ch>
.
Number of the channel to be noise figure calibrated.
The subsequent noise figure calibration commands must use the
same channel number.
Setting parameters:
<arg0> Number of the physical VNA port connected to the DUT output.
<arg1> Number of the physical VNA port connected to the DUT input.
<arg2> ON | OFF - Boolean parameter, indicating whether an external
attenuator is used.
<arg3> Attenuation of the source level during the source noise calibra-
tion.
<arg4> Attenuation of the source level during the measurement (i.e.
after calibration).
Example: See Chapter 8.2, "Condensed programming examples",
on page 1599.
Usage: Setting only
Options: R&S ZNA-K30
[SENSe<Ch>:]CORRection:COLLect:NFIGure[:ACQuire] <arg0>
Starts one of the three noise calibration sweeps that are required for a noise figure cali-
bration of the related channel.
The noise figure calibration is initiated using [SENSe<Ch>:]CORRection:COLLect:
NFIGure:STARt and must be terminated by one of the commands [SENSe<Ch>:
]CORRection:COLLect:NFIGure:END or [SENSe<Ch>:]CORRection:
COLLect:NFIGure:SAVE.
Note: A noise figure calibration must be terminated with one of the following com-
mands, to be sent immediately after the sequence of ...[:ACQuire] commands:
Abort the noise figure calibration, discarding the acquired calibration data.
Terminate the noise figure calibration, applying the acquired calibration data (the
active channel is calibrated) These commands ensure that the analyzer is able to
start a new measurement sweep.