EasyManua.ls Logo

R&S ZNA26 - Page 1353

R&S ZNA26
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1353User Manual 1178.6462.02 ─ 20
Parameters:
<FrequencyLimit> Frequency limit at which the noise figure sideband correction is
used.
Range: 10 MHz | 9 GHz
Default unit: Hz
Options: R&S ZNA-K30
Manual operation: See "Sideband Correction" on page 329
[SENSe<Ch>:]NFIGure:HARMonic[:MAXimumorder] <MaximumOrder>
Sets the sideband order of the sideband correction for a noise figure measurement in
channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<MaximumOrder> Sets the number of higher order sidebands which are measured
and corrected during calibration and measurement.
*RST: 3
Options: R&S ZNA-K30
Manual operation: See "Sideband Correction" on page 329
[SENSe<Ch>:]NFIGure:IFConst <IFFreq>
If manual selection is activated ([SENSe<Ch>:]NFIGure:IFConst:STATe ON, this
command allows you to select an IF for niose figure measurements in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<IFFreq> Default unit: Hz
Options: R&S ZNA-K30
Manual operation: See "General > IF Frequency" on page 329
[SENSe<Ch>:]NFIGure:IFConst:STATe <Boolean>
Allows you to switch between automatic and manual IF selection for niose figure mea-
surements in channel <Ch>
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> OFF (0)
Automatic selection
SCPI command reference