EasyManua.ls Logo

R&S ZNA26 - Page 1377

R&S ZNA26
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1377User Manual 1178.6462.02 ─ 20
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM OFF
Disable the measurement in the created sweep segment.
SEGM:COUN?
Query the number of segments.
Usage: Query only
Manual operation: See "Table Columns" on page 500
[SENSe<Ch>:]SEGMent<Seg>[:STATe]
<Boolean>
Activates or deactivates the sweep segment <Seg>. Sweep points belonging to inac-
tive segments only are not measured.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<Boolean> ON | OFF - activates or deactivates the measurement in sweep
segment <Seg>.
*RST: ON
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM OFF
Disable the measurement in the created sweep segment.
Manual operation: See "Table Columns" on page 500
[SENSe<Ch>:]SEGMent<Seg>:ADD
Inserts a new sweep segment using default channel settings ("Insert New Segment").
The added segment covers the frequency interval between the maximum frequency of
the existing sweep segments and the stop frequency of the entire sweep range.
Tip: Use [SENSe<Ch>:]SEGMent<Seg>:INSert to create a segment with specific
channel settings.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number. Segment numbers must be sequential.
If n segments exist already, the added segment must have the
segment number n+1.
SCPI command reference

Table of Contents

Other manuals for R&S ZNA26

Related product manuals