Command reference
R&S
®
ZNA
1417User Manual 1178.6462.02 ─ 20
*RST: The default frequency/frequency range corresponds
to the sweep range or CW frequency of the ana-
lyzer, i.e. <Numerator> = <Denominator> = 1, <Off-
set> = 0 Hz.
<SweepType> FB | CW
FB
Sweep with channel base frequency f
b
CW
Fixed frequency
Options: R&S ZNA-B8
Manual operation: See "Freq. Conversion" on page 616
SOURce<Ch>:RLO:PABSolut
<Mode>
Allows you to define the output power mode at the (optional) rear LO Out port for chan-
nel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Mode> OFF (0)
Use the channel base power P
b
ON (1)
Use a fixed value of 0 dBm
Options: R&S ZNA-B8
Manual operation: See "Channel Base Power" on page 575
SOURce<Ch>:RLO:PERMenable <Boolean>
Defines whether the source power at the (optional) rear LO Out port is permanently on
in channel <Ch> ("generator port") .
Suffix:
<Ch>
.
Chanel number
Parameters:
<Boolean> OFF (0)
Power is only on for partial measurements that require the port
as a drive port.
ON (1)
Power is permanently on
*RST: OFF
Options: R&S ZNA-B8
Manual operation: See "Source Gen" on page 615